Configuring Multiple Mode with a PX2 Light Source and USB2000+
On December 9, 2011 By Optics Ocean
When using the PX2 in multiple mode with a USB2000+ the integration time should be an integer multiple of the continuous strobe delay, and the integration time must be set before setting the continuous strobe delay.
Example Set the integration time to 100msec, Thenย Set the continuous strobe delay to 25msec (100 being an integer [...]